# Calibration Inspection

<p class="doc-lead">A Calibration Inspection is the record for one instrument. It holds the instrument details, the environmental conditions, the reference instruments used, the test results, and who calibrated it. Once the inspection is Completed and approved, the printed copy is the calibration certificate.</p>

<div class="doc-toc">
<p class="doc-toc-title">On this page</p>
<div class="doc-toc-section">
<span class="doc-toc-group-title">Getting started</span>
<div class="doc-toc-links">
<a href="#/calibration/calibration-inspection?id=overview">Overview &amp; List View</a>
<a href="#/calibration/calibration-inspection?id=quick-start">Quick Start</a>
</div>
</div>
<div class="doc-toc-section">
<span class="doc-toc-group-title">Form reference</span>
<div class="doc-toc-links">
<a href="#/calibration/calibration-inspection?id=certificate-section">Certificate</a>
<a href="#/calibration/calibration-inspection?id=unit-under-calibration">Unit Under Calibration</a>
<a href="#/calibration/calibration-inspection?id=calibration-details">Calibration Details</a>
<a href="#/calibration/calibration-inspection?id=reference-instruments">Reference Instruments</a>
<a href="#/calibration/calibration-inspection?id=results-and-sign-off">Test Results and Sign Off</a>
</div>
</div>
<div class="doc-toc-section">
<span class="doc-toc-group-title">View and actions</span>
<div class="doc-toc-links">
<a href="#/calibration/calibration-inspection?id=view">View</a>
<a href="#/calibration/calibration-inspection?id=statuses">Statuses</a>
<a href="#/calibration/calibration-inspection?id=change-status">Change Status</a>
<a href="#/calibration/calibration-inspection?id=actions">Actions</a>
<a href="#/calibration/calibration-inspection?id=certificate">Certificate Print</a>
</div>
</div>
<div class="doc-toc-section">
<span class="doc-toc-group-title">More</span>
<div class="doc-toc-links">
<a href="#/calibration/calibration-inspection?id=best-practices">Best Practices</a>
<a href="#/calibration/calibration-inspection?id=faq">FAQ</a>
<a href="#/calibration/calibration-inspection?id=related">Related</a>
</div>
</div>
</div>

<hr class="doc-section-divider">

<h2 id="overview">Overview &amp; List View</h2>

Go to <b>Calibration → Calibration Inspection</b> from the left menu. The list is the register of all inspections and certificates. Click a row to open it.

<b>List columns</b>

`Id`, `Certificate No.`, `CRN #`, `Customer`, `Calibrated Item`, `Serial No.`, `Certificate Type`, `Date of Issue`, `Inspection Status`, `Status`

<b>Quick filters</b>

| **Filter** | **Shows** |
|------------|-----------|
| Default | All inspections except Cancelled |
| Pending | Inspection status Pending |
| In Progress | Inspection status In Progress |
| On Hold | Inspection status On Hold |
| Completed | Inspection status Completed |
| Cancelled | Cancelled inspections |

<b>Advanced filters</b>

`Id`, `Certificate No.`, `Calibration Request Note`, `Serial No.`, `Certificate Type`, `Inspection Status`, `Status`, `Date of Issue`, `Created date`

There is no <b>New</b> button on the list. Inspections are created from the Calibration Request Note.

<hr class="doc-section-divider">

<h2 id="quick-start">Quick Start — Create an Inspection and Issue a Certificate</h2>

<ol>
<li>Open an <b>Approved</b> <a href="#/calibration/calibration-request-note">Calibration Request Note</a>.</li>
<li>On the instrument line, click <b>+ Create Inspection</b>. The form opens in a window with the customer and instrument details already filled.</li>
<li>Check the <b>Certificate</b>, <b>Unit Under Calibration</b>, and <b>Calibration Details</b> sections. Fill the environmental conditions.</li>
<li>Add the reference instruments you used.</li>
<li>Write the test results and remarks, and choose <b>Calibrated By</b>.</li>
<li>Click <b>Save</b>. The inspection opens in view.</li>
<li>Set <b>Inspection Status</b> to <b>Completed</b> using <b>Change Status</b>, then click <b>Approve</b>.</li>
<li>Click <b>Print</b> to print the certificate.</li>
</ol>

Every field marked * is mandatory. <b>Save And Approve</b> saves and approves in one step.

<hr class="doc-section-divider">

<h2 id="certificate-section">Certificate</h2>

| **Field** | **Description** |
|-----------|-----------------|
| Calibration Request Note | The note this inspection belongs to. Read only. |
| Customer | Customer on the note. Read only. |
| Certificate No. * | Hidden when auto-generated numbering is on for Calibration Inspection. See <a href="#/settings/number-settings">Number Settings</a>. |
| Date of Issue * | Date the certificate is issued. Defaults to today. |
| Certificate Type * | Calibration Certificate, Test Certificate, or Defect / Reject Certificate. Sets the title printed at the top. |
| Inspection Status * | Pending, In Progress, On Hold, or Completed. Defaults to Pending. |

<hr class="doc-section-divider">

<h2 id="unit-under-calibration">Unit Under Calibration</h2>

Filled from the instrument line on the request note. You can change any of it.

| **Field** | **Description** |
|-----------|-----------------|
| Calibrated Item * | Name of the instrument. Defaults to the service item name. |
| Make | From the request note. |
| Model | From the request note. |
| Serial No. | The instrument's serial number from the request note. |
| ID No. | The customer's or lab's identification number. |
| Calibrating Range | Range of the instrument being calibrated. |
| Resolution | Resolution of the instrument. |

<hr class="doc-section-divider">

<h2 id="calibration-details">Calibration Details</h2>

| **Field** | **Description** |
|-----------|-----------------|
| Received Date * | Defaults to the request note's receipt date. |
| Calibration Date * | Defaults to today. |
| Recommended Due Date | Filled automatically as the calibration date plus the request note's frequency, minus one day. For example, 12 months from 1 Mar 2026 gives 28 Feb 2027. You can overwrite it. |
| Calibration Performed At | Place where the calibration was done. |
| Calibration Procedure * | Procedure reference, SOP/001 to SOP/010. |
| Temperature | Ambient temperature during the calibration, in °C. |
| Humidity | Relative humidity, in %. |
| Atm. Pressure | Atmospheric pressure, in mbar. |

<hr class="doc-section-divider">

<h2 id="reference-instruments">Reference Instruments</h2>

List the instruments you calibrated against. Add one row for each. The serial number column fills automatically.

| **Field** | **Description** |
|-----------|-----------------|
| Description * | A reference instrument from <a href="#/calibration/reference-instrument">Reference Instrument</a>. Only active instruments are offered. Use the add button to create one without leaving the form. |
| Serial / ID No. | Filled from the reference instrument. |
| Validity | Filled from the reference instrument. |
| Traceability | Filled from the reference instrument. |

Change the details on the Reference Instrument master, not on the inspection. If a reference instrument is edited later, certificates show its current details.

<hr class="doc-section-divider">

<h2 id="results-and-sign-off">Test Results and Sign Off</h2>

| **Field** | **Description** |
|-----------|-----------------|
| Test Results | Rich-text area for the readings. Use the <b>Insert table</b> tool in the editor to lay out readings in a table. Printed centred. |
| Remarks | Rich-text area printed as <b>Condition Of Measurement &amp; Remarks</b>. Method notes, tolerance statements, and the like. |
| Calibrated By * | The employee who did the calibration. Printed as the signature. Active employees only. |
| Internal Notes | Notes for staff. Not printed on the certificate. |

<hr class="doc-section-divider">

<h2 id="view">View</h2>

The view shows the certificate in the standard document layout with the company header.

<ul>
<li><b>Title:</b> set by Certificate Type — CALIBRATION CERTIFICATE, TEST CERTIFICATE, or DEFECT / REJECT CERTIFICATE.</li>
<li><b>Header:</b> Certificate No. and Date of Issue. On screen the view also shows the Request Note (click for a quick view) and the Inspection Status.</li>
<li><b>Customer Details:</b> the customer name and address. On screen, click it for the customer quick view.</li>
<li><b>Unit Under Calibration - Specifications:</b> instrument details on the left, dates on the right, the Environmental Conditions block, and the Calibration Procedure in bold below.</li>
<li><b>Details of Reference Instruments Used for Calibration:</b> description, serial no. / ID no., validity, and traceability.</li>
<li><b>Test Results</b> and <b>Condition Of Measurement &amp; Remarks</b>.</li>
<li><b>Signature:</b> Calibrated By, then the disclaimer and the line <i>END OF CERTIFICATE</i>.</li>
</ul>

<hr class="doc-section-divider">

<h2 id="statuses">Statuses</h2>

Each inspection has two statuses.

<b>Document status</b>

| **Status** | **Meaning** |
|------------|-------------|
| Draft | Saved, not approved. Edit and Delete are available. |
| Approved | Approved. This is the certificate of record. |
| Cancelled | Stopped. Excluded from the default list. The instrument line can be inspected again. |

<b>Inspection status</b>

| **Status** | **Meaning** |
|------------|-------------|
| Pending | Not started. The default. |
| In Progress | Calibration under way. |
| On Hold | Paused, for example waiting for the customer. |
| Completed | Calibration finished. Required before approval. |

When every instrument on the request note has a Completed and approved inspection, the <a href="#/calibration/calibration-request-note?id=auto-close">request note closes automatically</a>.

<hr class="doc-section-divider">

<h2 id="change-status">Change Status</h2>

Open the more menu (⋮) and choose <b>Change Status</b>. A window shows the <b>Certificate No.</b> and <b>Current Status</b>. Choose the <b>New Status</b> and click <b>Update</b>.

Change Status works on Draft, Approved, and For-Revisal inspections, so you can mark an approved inspection as On Hold or move it back to In Progress. It does not work on a Cancelled inspection.

<hr class="doc-section-divider">

<h2 id="actions">Actions</h2>

<h3>List</h3>

| **Action** | **When** |
|------------|----------|
| View | One row. |
| Edit | Draft. |
| Delete | Draft. |

<h3>View toolbar and more menu</h3>

| **Action** | **When** | **Description** |
|------------|----------|-----------------|
| Approve | Draft and Inspection Status is Completed | Approves the inspection. The button stays hidden until the status is Completed. It appears as soon as you change the status, without refreshing. |
| Edit | Draft | Open the full edit form. |
| Delete | Draft | Remove the inspection. Frees the instrument line. |
| Print | Any | Print the certificate. |
| Change Status | Draft, Approved | Change the inspection status. |
| Add/Update Note | Any | Add or change the internal note. |
| Cancel | Approved | Cancels the inspection and frees the instrument line. |

<hr class="doc-section-divider">

<h2 id="certificate">Certificate Print</h2>

Click <b>Print</b> on the view. The print layout is the calibration certificate: the title for the certificate type, the customer table, the specifications grid with Environmental Conditions, the reference instruments, test results, remarks, the Calibrated By signature, the disclaimer, and the end-of-certificate line. Fields marked as screen-only — the Request Note and the Inspection Status — are not printed.

Check the print preview before issuing. Fill the optional fields (ID No., Calibrating Range, Resolution, and the environmental conditions) so the printed grid has no blanks.

<hr class="doc-section-divider">

<h2 id="best-practices">Best Practices</h2>

<ul>
<li>Add every reference instrument to the <a href="#/calibration/reference-instrument">Reference Instrument</a> master first, with validity and traceability, so certificates always show correct details.</li>
<li>Use <b>In Progress</b> and <b>On Hold</b> to track work. The <a href="#/calibration/calibration-dashboard">dashboard</a> counts them.</li>
<li>Do not use a reference instrument whose validity has expired. The dashboard flags these.</li>
<li>Complete the test results before setting the status to Completed. Approved inspections cannot be edited.</li>
<li>Check the print preview before the first certificate of a new instrument type.</li>
</ul>

<hr class="doc-section-divider">

<h2 id="faq">FAQ</h2>

<b>Where is the Approve button?</b><br>
Approve shows only on a Draft inspection whose Inspection Status is <b>Completed</b>. Use <b>Change Status</b> first.

<b>Can I create an inspection from the Calibration Inspection list?</b><br>
No. Create it from an instrument line on an approved Calibration Request Note.

<b>Why can't I create a second inspection for an instrument?</b><br>
Each instrument line allows one active inspection. Cancel (or delete a Draft) first.

<b>Can I edit an approved inspection?</b><br>
No. Edit and Delete are for Draft inspections only. Cancel the approved inspection and create a new one for the instrument line.

<b>Why does the request note show a different status from the inspection?</b><br>
The note's Inspection Status column shows the inspection status while the inspection is approved, and the document status otherwise.

<b>Does the recommended due date follow the request note?</b><br>
Yes. It is calibration date plus the note's frequency (3, 6, or 12 months) minus one day, and it recalculates when you change the calibration date.

<hr class="doc-section-divider">

<h2 id="related">Related</h2>

<ul>
<li><a href="#/calibration/">Calibration Module</a></li>
<li><a href="#/calibration/calibration-request-note">Calibration Request Note</a></li>
<li><a href="#/calibration/reference-instrument">Reference Instrument</a></li>
<li><a href="#/calibration/calibration-dashboard">Calibration Dashboard</a></li>
<li><a href="#/settings/number-settings">Number Settings</a></li>
</ul>

<section class="ai-index-only">

## Errors and warnings

Exact messages for Calibration Inspection from the ERP screen and `fs_api`. Visitors do not see this section; Vertex AI Search uses it to answer questions about on-screen errors.

### Approve before completed

- **Type:** Error
- **Action:** Approve, Save And Approve
- **When:** Inspection Status is not Completed.
- **Example message:** *Inspection must be completed before approval*
- **Outcome:** The inspection stays Draft.
- **Solution:** Change Inspection Status to Completed, then approve.

### Inspection already exists for the line

- **Type:** Error
- **Action:** Create Inspection, Save, Save And Approve, Update
- **When:** Another inspection that is not cancelled exists for the same instrument line.
- **Example message:** *Inspection already exists for this calibration request line*
- **Outcome:** The inspection is not saved.
- **Solution:** Open the existing inspection, or cancel it and create a new one.

### Request note not approved

- **Type:** Error
- **Action:** Create Inspection, Save, Save And Approve
- **When:** The Calibration Request Note is Draft or Cancelled.
- **Example message:** *Calibration request is not approved*
- **Outcome:** The inspection is not created.
- **Solution:** Approve the request note first.

### Missing or invalid request line

- **Type:** Error
- **Action:** Save, Save And Approve
- **When:** The inspection is not linked to an instrument line, or the line does not belong to the note.
- **Example message:** *Calibration request line is required* or *Invalid calibration request line*
- **Outcome:** The inspection is not saved.
- **Solution:** Create the inspection from the instrument line on the request note.

### Invalid inspection status

- **Type:** Error
- **Action:** Change Status
- **When:** The new status is not Pending, In Progress, On Hold, or Completed.
- **Example message:** *Invalid inspection status*
- **Outcome:** The status is not changed.
- **Solution:** Choose one of the listed statuses.

### Change status on a cancelled inspection

- **Type:** Error
- **Action:** Change Status
- **When:** The inspection is Cancelled.
- **Example message:** *Inspection is cancelled*
- **Outcome:** The status is not changed.
- **Solution:** Create a new inspection for the instrument line.

### Future or back-dated date

- **Type:** Error (blocks save)
- **Action:** Save, Save And Approve
- **When:** Date of Issue is in the future and future-date transactions are off, or the user cannot enter previous dates, or the date is before the allowed cutoff.
- **Example message:** *Future date transactions are not allowed.* or *Previous date transaction entries are not allowed.* or *Transaction date cannot be before YYYY-MM-DD.*
- **Outcome:** The inspection is not saved.
- **Solution:** Set Date of Issue to an allowed date, then save.

### Edit or delete not allowed

- **Type:** Error
- **Action:** Edit, Delete
- **When:** The inspection is not Draft.
- **Example message:** *Edit not allowed in the current status* or *Invalid status*
- **Outcome:** The inspection is not changed.
- **Solution:** Edit and delete Draft inspections only. Cancel an approved inspection.

</section>